2. Right-click an existing instrument parameter file listed in
3. Enter a new file name (e.g., "Instrument parameters
4. Go to the Web page cited above and open the file
5. Enter the dispensing pressures and pulse time into the
6. Save the new instrument parameter file.
7. When creating a new Run Setup, select this new
PyroMark Q96 ID User Manual 01/2016
the upper pane and select "Duplicate" from the context
menu.
Code 0005").
"Managing instrument parameters for the PyroMark Q96
ID", found under the User Support tab. In the provided
table, find the dispensing pressures and pulse time
settings that correspond to the Code number printed on
the PyroMark Q96 Cartridge to be used.
corresponding fields of the "Instrument Parameters"
dialog box.
instrument parameter file from the list of instrument
parameter files.
Operating Procedures
5-5