Maintenance
7.5

Changing the fuses

The fuses of the PyroMark Q96 ID should be checked if the
instrument power indicator is not illuminated, even though
the power cable is connected to the mains power outlet and
the standby switch is on.
Important: Ensure that the PyroMark Q96 ID is disconnected
from the mains power before removing fuses from the
instrument.
Have a qualified technician (e.g., facilities staff) ensure that
the mains power outlet is live.
The following item is required:
Procedure:
1. Turn off the standby switch at the front of the instrument
2. If necessary, carefully turn the instrument, so that the rear
3. Use a small screwdriver to release the securing lugs on
4. Pull out the fuse holder using your fingers. The fuse
5. Insert the fuse holder (it is keyed and can only be
6. Reposition the instrument, reconnect the power cable
7. If the instrument still does not work, please contact
7-4
Small flat-head screwdriver
and disconnect the mains power cable.
panel can be easily accessed.
each side of the fuse holder.
holder contains two fuses (type F 6.3 A H, 250 V). Check
both fuses and replace as necessary.
Note: Replace fuses only with fuses of the specified type
and current rating.
replaced in one way) and push it into place using your
fingers.
and turn the standby switch on.
QIAGEN Technical Services.
PyroMark Q96 ID User Manual 01/2016