Prodigit 3270 Series Operation Manual page 75

Ac/dc high power electronic load
Table of Contents

Advertisement

66 PRODIGIT
• The INRUS parameters setting:
Pressing the Item key once will cause the button to illuminate. The message
"INRUS PRESS START" will be shown across the displays.
Each press of the setting button moves the menu on one step. The Left and right
LCDs show the currently selected test parameter as text. The value is adjusted by
The rotary knob and can be read from the Right display during Setting.
The setting sequence is shown below:
INRUS FREQ: setting the INRUS FREQ, the Left 5 digit monitor display the
"INRUS", the right upper 5 digit monitor display the "FREQ", and Right lower
monitor display setting value, the unit is "Hz", use the knob and button to set the
Range from DC and 40~ 440Hz.
INRUS ISTAR : setting the INRUS ISTAR, the Left 5 digit monitor display the
"INRUS" ,the right upper 5 digit monitor display the "ISTAR", and right lower
monitor display setting value, the unit is "A". Use the knob and button to set the
starting current value, the setting range from 0.000 A to 75.000A.
INRUS PRESS START
INRUS FREQ
INRUS ISTAR
INRUS ISTEP
INRUS TIME
INRUS ISTOP
.