3-7.

Protection features

The protection features of the 3270 series Electronic load modules are as follows:
3.7.1. Overvoltage protection: The Electronic Load input will turn OFF if the overvoltage
circuit is tripped. The message OVP will be displayed on the LCD. When the OVP
fault has been removed the load can be set to sink power again. While the unit
will attempt to protect itself given an OVP state it is strongly advised to guard
against any potential OVP fault state by using external protection and the correctly
rated electronic load.
The Overvoltage protection circuit is set at a predetermined voltage and cannot be
adjusted. The OVP level is 105% of the 3270 Series nominal voltage rating.
CAUTION: Never apply an AC voltage to the input of the 3270 series Load. Do
not apply a DC voltage that is higher than 3270 series Load rating. If this
advice is ignored it is likely that damage will be caused to the electronic load
module. This damage will not be covered by the warranty.
3.7.2. Over current protection (OCP): The OCP protection will engage if the current being
taken by the load reaches 105% of the load module's maximum current. The
message OCP will be displayed on the front panel and the unit will switch to its
LOAD OFF state. Once the source of the over current has been removed the load
can be switched on again.
3.7.3. Over power protection (OPP): The 3270 series Electronic Load monitors the power
dissipation level. The input to the load is automatically switched to LOAD OFF if the
power dissipation is greater than 105% of the rated power input. If an over power
condition occurs the display will show OPP
3.7.4. Over temperature protection (OTP): The load internal temperature at the heat sink is
monitored. If the temperature reaches approximately 100°C the OTP message will
be displayed and the unit will automatically switch to the LOAD OFF state. If an
OTP error occurs please check the ambient temperature is between 0 to 40°C. Also
ensure that the front and rear air vents of the mainframe are not obstructed. The air
flow is taken from the front of the mainframe and exhausted from the rear.
Therefore a suitable gap needs to be left at the rear of the mainframe. A minimum
of 15cm is recommended. After a suitable cooling period the load can be switched.
3270 Series Operation Manual 91