108 PRODIGIT
OPP: STOP
Syntax: [PRESet:] OPP: STOP {SP}{NR2}{;NL}
Purpose:Set and read the maximum value of OPP test
Description:This command is used for setting the maximum value (P-STOP)of OPP test
STIME
Syntax: [PRESet:] STIME {SP}{NR2}{;NL}
Purpose:Set and read time of the short-circuit test
Description:This command is used for setting time of the short-circuit test. If time set to
0, it means that have no the time limit and continue to be short –circuited. The unit is
milli-second (ms)
PF
Syntax: [PRESet:] PF {SP}{+-}{NR2}{;NL}
Purpose:Set and read Power factor.
Description:This command is set Power factor, the setting range is 0.01 ~ 1.00.
CF
Syntax: [PRESet:] CF {SP}{NR2}{;NL}
Purpose: Set and read the crest factor.
Description:This command is set crest factor,the setting range is 1.4142~5.0.
BATT: MODE
Syntax: [PRESet:] BATT:MODE {SP}{CCCRCVCPLIN}{;NL}
Purpose:Set and read the Battery test mode.
Description:This command is set and read the Battery test mode.
BATT: TIME
Syntax:[PRESet:] BATT:TIME {SP} }{NR1 }{;NL}
Purpose:Set and read the Battery test time.
Description:This command is set and read the Battery test time, the setting range is
DISC: TIME
Syntax: [PRESet:] DISC: TIME {?}{;NL}
Purpose:Read the battery discharge time.
Description:This command is when the test end, read the battery discharge time, the
DISC: AH
Syntax: [PRESet:] DISC: AH {?}{;NL}
Purpose: Read the battery capacity.
Description:This command is when the test end, read the battery capacity.
[PRESet:] OPP: STOP {?} {;NL}
[PRESet:] STIME {?} {;NL}
[PRESet:] PF {?} {;NL}
[PRESet:] CF {?} {;NL}
[PRESet:] BATT: MODE {?}{;NL}
[PRESet:] BATT: TIME {?}{;NL}
1S~99999S.
Range of 1S ~ 99999S.