Prodigit 3270 Series Operation Manual page 115

Ac/dc high power electronic load
Table of Contents

Advertisement

106 PRODIGIT
CVIVOLT :{ AB}
Syntax: [ PRESet:] CVI: {AB} {SP}{ NR2}{;NL}
Purpose:Set and read the value of Voltage
Description:This command is for setting the required value of Voltage, and the unit is V.
TCONFIG
Syntax: [PRESet:] TONFIG {NORMAL|OCP|OVP|OPP|SHORT|NLIN|NLCR|FUSE
Purpose: Set and read the of test Item
Description: There are nine options of this command. Those are NORMAL mode, OCP
test, OPP test, SHORT, NLIN, NLCR, FUSE, BATT, TRANS, INRUSH, SURGE test.
ITIME
Syntax: [PRESet:] TIME {SP}{NR2}{;NL}
Purpose: Set and read the INRUSH current time
Description: Use this command to set the interval for current decrement. The setting
Range is 0.1ms~100.0ms.
ISTART
Syntax: [PRESet:] ISTART {SP}{NR2}{;NL}
Purpose: Set and read the starting current set point for the inrush current test.
Description: The starting current is set to twice the current specification.
ISTEP
Syntax: [PRESet:] ISTEP {SP}{NR2}{;NL}
Purpose: Set and read the set value of the decrement current of the inrush current test.
Description: The step current is set to twice the current specification.
ISTOP
Syntax: [PRESet:] ISTOP {SP}{NR2}{;NL}
Purpose: Set and read the set value of the minimum current for the inrush current test.
Description: Minimum current setting range current specification.
SURGE:Tn
Syntax: [PRESet:]SURGE:Tn {SP}{NR2}{;NL}
Purpose: Set and read the time setting for the surge current test.
Description: n: 1~3, the time to load current in three stages.
When n=1, 2, the time setting range is 0.01~0.50 seconds.
When n=3, the time setting range is 0.01~9.99 seconds or continuous loading.
[ PRESet:] CVI: {AB} {?} {;NL}
|BATT|TRANS|INRUSH|SURGE}{;NL}
[PRESet:] TONFIG {?} {;NL}
[PRESet:]ITIME {?}
[PRESet:] ISTART {?}
[PRESet:] ISTEP {?}
[PRESet:] ISTOP {?}
[PRESet:] SURGE:Tn {?}