Prodigit 3270 Series Operation Manual page 118

Ac/dc high power electronic load
Table of Contents

Advertisement

EXTIN: ON/OFF
*( this function is optional.)
Syntax:[PRESet:] EXTIN: {SP} ONOFF}{;NL}
[PRESet:] EXTIN{?}{;NL}
Purpose: Set the external input signal .
Description:This command is set EXTIN ON or OFF.
TURBO: {SP}{ONOFF}
Syntax:[PRESet:] TURBO{ONOFF}{;NL}
[PRESet:] TURBO {?}{;NL}
Purpose: Set and read theTURBO mode can be set to ON or OFF.
Description:In TURBO mode, output double maximum rated current in short time.
FUSE:CC
Syntax:[PRESet:] FUSE:CC {SP}{NR2 }{;NL}
[PRESet:] FUSE: CC {?}{;NL}
Purpose: Set and read fuse test current value.
Description:This command is to set or read the fuse test current value, In normal mode
The range is 0 ~ 37.5A, In TURBO mode the range is 0 ~ 75A.
FUSE: TIME
Syntax: [PRESet:] FUSE: TIME {SP} }{NR2 }{;NL}
[PRESet:] FUSE: TIME {?}{;NL}
Purpose: Set and read fuse test time.
Description:This command is to set or read the fuse test time, the setting range is 0.1S
~ 9999.9S.
FUSE: TYPE
Syntax: [PRESet:] FUSE:TYPE {SP} }{TRIPNTRIP }{;NL}
[PRESet:] FUSE: TYPE {?}{;NL}
Purpose: Set and read fuse type.
Description:This command is to set or read fuse TRIP or NTRIP.
FUSE: REP
Syntax:[PRESet:] FUSE:REP {SP} }{NR1 }{;NL}
[PRESet:] FUSE: REP {?}{;NL}
Purpose: Set and read the fuse repeat tests number of times.
Description:Set and read the fuse repeat tests number of times,the setting range is 0 ~
255 times.
TRIP: TIME
Syntax: [PRESet:] TRIP: TIME {?}{;NL}
Purpose: read the fuse fusing time.
Description:This command is when the test end, read the fuse fusing time.
3270 Series Operation Manual 109