Prodigit 3270 Series Operation Manual page 144

Ac/dc high power electronic load
Table of Contents

Advertisement

5.13.1.8. Press START/STOP Test key.
5.13.1.9. The UUT's output voltage drop-out lower than the threshold voltage (V-th
Setting), and the OPP trip point is between W_Hi and W_Lo limitation,
Then Right upper 5 digits LCD display will shows "PASS", otherwise
Shows "FAIL".
5.13.2. Remote control OPP
EX :
REMOTE
TCONFIG OPP
OPP:START 3
OPP:STEP 1
OPP:STOP 5
VTH 3.0
WL 0
WH 5
NGENABLE ON
START
TESTING?
NG?
OPP?
STOP
3270 Series Operation Manual 135
(Set Remote)
(Set OCP test)
(Set start load watt 3W)
(Set step load watt 1W)
(Set stop load watt 5W)
(Set OPP VTH 3.0V)
(Set watt low limit 0W)
(Set watt high limit 5W)
(Set NG Enable ON)
(Start OPP testing)
(Ask Testing? 1:Testing,0:Testing End)
(Ask PASS/FAIL?,0:PASS,1:FAIL)
(Ask OPP watt value)
(Stop OPP testing)