42 PRODIGIT
Press the CF key, setting range from 1.4 to 5.0, step 0.1 by rotating the Setting
Knob, press the ENTER key after the completion of the setting will be automatically
Stored.
Note: CF( crest factor) range 1.4 ~ 5.0, 3270 series full scale current is 3 times the
peak, if use the CF peak 5.0, 3270 full scale current so the current must be reduced
to 22.5A, in order to reach the peak 5.0.
Current phase modulation waveform load
90 degree SCR Trailing edge current waveform
90 degree SCR Leading edge current waveform