Prodigit 3270 Series Operation Manual page 119

Ac/dc high power electronic load
Table of Contents

Advertisement

110 PRODIGIT
TRANS: TIME
Syntax: [PRESet:] TRANS: TIME {?}{;NL}
Purpose: read UPS Transfer time.
Description:This command is when the test end, read the UPS Transfer time.
AVG
Syntax:
Purpose: Set and read back the average 1, 2, 4, 8, and 16.
Description: Set and read back the average 1, 2, 4, 8, and 16, the default is 1 without
CPRSP
Syntax:
Purpose: Set and read back the CPRSP 0~7. the default is 0.
Description: CPRSP is set to the constant power response speed 0~4 for linear current
constant power load, 0 is the fastest to adjust the load power response, 3 is the slowest
4~7 is the standard current constant power load 4 to adjust the load power The
response is the fastest, and the slowest default is 0.
CYCLE
Syntax:
Purpose: Set and read back the CYCLE ,can be set from 1 to 16.
Description: The set is 8, that is 8 weeks to do the meter value processing.
BW
Syntax: [PRESet:] BW {SP} {NR2}{;|NL}
Purpose: Set and read the BW 0~15.
Description: Set and read the bandwidth from 0 to 15 bandwidth, 15 is the fastest, and
FREQ
Syntax: [PRESet:] FREQ {SP} {NR2}{;|NL}
Purpose: Set and read the frequency .
Description: Set and read the frequency ,range from 40 to 440 Hz.
REP: COUNT
Syntax: [PRESet:] REP: COUNT? {;|NL}
Purpose: use fuse test.
Description: Read the number of repeated tests.
[PRESet:] AVG {SP} {NR2}{;|NL}
[PRESet:] AVG? {; |NL}
averaging.
[PRESet:] CPRSP {SP} {NR2}{;|NL}
[PRESet:] CPRSP? {;|NL}
[PRESet:] CYCLE {SP} {NR2}{;|NL}
[PRESet:] CYCLE? {;|NL}
[PRESet:] BW? {;|NL}
the initial Value is 13.
[PRESet:] FREQ? {;|NL}