52 PRODIGIT
"OPP PRESS START" will be shown across the displays.
Each press of the Setting button moves the menu on one step. The Left and Middle
LCDs show the currently selected test parameter as text. The value is adjusted by
The rotary knob and can be read from the Right display during Setting.
The setting sequence is shown below:
The right upper 5 digit monitor display the Turbo and right lower monitor display
"OFF" ,use the knob and the key to switch ON or OFF.
PSTAR: setting the start power, the Left 5 digit monitor display the "OPP" ,the right
upper 5 digit monitor display the "PSTAR", and right lower monitor display setting
value, the unit is "W". The range is 0.1W to the full scale of the CP mode
specification.
PSTEP: setting the increment step power, the Left 5 digit monitor display the
"OPP" ,the right upper 5 digit monitor display the "PSTEP", and right lower monitor
display setting value, the unit is "W". The range is 0.1W to the full scale of the CP
mode specification.
OPP PRESS START
OPP TURBO
OPP PSTAR
OPP PSTEP
OPP PSTOP
OPP VTH