34 PRODIGIT
Setting Upper limit current IH , the right upper 5 digit monitor display "I-Hi"
and right lower monitor display upper limit of the voltmeter with the unit as
"A", The I-Hi set range from 0.000 A to 40.000A step 0.001A by rotating the
Setting knob.
Setting lower limit current IL , the right upper 5 digit monitor display "I-Lo" and
right lower monitor display lower limit of the voltmeter with the unit as
"A" ,The I-Lo set range from 0.000 A to 40.000A step 0.001A by rotating
the Setting knob.
Setting Upper limit power WH, the right upper 5 digit monitor display "W-Hi"
and right lower monitor display upper limit of the voltmeter with the unit as
"W", The W-Hi set range from 0 W to 4000.0W step 1W by rotating the
Setting knob.
Setting lower limit power WL, the right upper 5 digit monitor display "W-Lo"
and right lower monitor display lower limit of the voltmeter with the unit as
"W", The W-Lo set range from 0.0 W to 4000.0W step 0.1W by rotating the
Setting knob.
Setting Upper limit power VAH, the right upper 5 digit monitor display "VA-Hi"
and right lower monitor display upper limit of the voltmeter with the unit as
"VA", The VA-Hi set range from 0 VA to 4000.0VA step 0.1VA by rotating
the Setting knob.
Setting lower limit power VAL, the right upper 5 digit monitor display "VA-Lo"
and right lower monitor display lower limit of the voltmeter with the unit as
"W", The VA-Lo set range from 0.0 VA to 4000.0VA step 0.1VA by rotating
the Setting knob.