SURGE:Sn
Syntax: [PRESet:]SURGE:Sn {SP}{NR2}{;NL}
[PRESet:] SURGE:Sn {?}
Purpose: Set and read the load current value of the surge current test.
Description: n: 1~3, the load current in three stages.
When n=1, 2, the load current setting range is twice the current specification.
When n=3, the load current setting range is the current specification.
OCP: START
Syntax: [PRESet:] OCP: START {SP}{NR2}{;NL}
[PRESet:] OCP: START {?} {;NL}
Purpose:Set and read the initial value of OCP test
Description:This command is used for setting the required initial value (I-START) of
OCP test
OCP: STEP
Syntax: [PRESet:] OCP:STEP {SP}{NR2}{;NL}
[PRESet:] OCP:STEP{?} {;NL}
Purpose:Set and read the increasing value of OCP test
Description:This command is used for setting the increasing value(I-STEP) of OCP test
OCP: STOP
Syntax: [PRESet:] OCP:STOP {SP}{NR2}{;NL}
[PRESet:] OCP:STOP {?} {;NL}
Purpose:Set and read the maximum value of OCP test
Description:This command is used for setting the maximum value (I-STOP) of OCP
test.
VTH
Syntax: [PRESet:] VTH {SP}{NR2}{;NL}
[PRESet:] VTH {?} {;NL}
Purpose:Set and read the value of the Threshold Voltage
Description : This command is used for setting the Threshold Voltage. That is the
OCP/OPP of this Load model when the output voltage of appliance is lower
or equaled to the VTH
OPP: START
Syntax: [PRESet:] OPP: START {SP}{NR2}{;NL}
[PRESet:] OPP: START {?} {;NL}
Purpose:Set and read the initial value of OPP test
Description:This command is used for setting the initial value (P-START) of OPP Test
OPP: STEP
Syntax: [PRESet:] OPP:STEP {SP}{NR2}{;NL}
[PRESet:] OPP:STEP {?} {;NL}
Purpose:Set and read the increasing value of OPP test
Description:This command is used for setting the increasing value (P-STEP)of OPP
Test