NL+CR CR: setting the NL+CR CR resistance point, the Left 5 digit monitor display
the "NL+CR" ,the right upper 5 digit monitor display the "CR", and right lower
monitor display setting value, the unit is "Ω" , use the knob and button to set the
CR value from 1.6000Ω to the full scale of the CR mode specification.
The FUSE parameters setting:
• Pressing the Item key once will cause the button to illuminate. The message
"FUSE PRESS START" will be shown across the displays.
• Each press of the setting button moves the menu on one step. The Left and right
LCDs show the currently selected test parameter as text. The value is adjusted by
The rotary knob and can be read from the Right display during Setting.
The setting sequence is shown below:
FUSE PRESS START
FUSE TURBO OFF
FUSE CC
FUSE TIME
FUSE REP.
FUSE TYPE TRIP
Setting the fuse TURBO, The Left 5 digit monitor display the "FUSE", the Right
Upper 5 Digit monitor display the "TURBO", and right lower monitor Display OFF;
Use the knob and the key to ON or OFF.
3270 Series Operation Manual 57