5.12.1.8. Press START/STOP test key.
5.12.1.9. The UUT's output voltage drop-out lower than the threshold voltage (V-th
Setting), and the OCP trip point is between I_Hi and I_Lo limitation, then
Right upper 5 digits LCD display will shows "PASS", otherwise shows "FAIL".
5.12.2. Remote control OCP
EX
REMOTE
TCONFIG OCP
OCP:START 0.1
OCP:STEP 0.01
OCP:STOP 2
VTH 3.0
IL 0
IH 2
NGENABLE ON
START
TESTING?
NG?
OCP?
STOP
(Set Remote)
(Set OCP test)
(Set start load current 0.1A)
(Set step load current 0.01A)
(Set stop load current 2A)
(Set OCP VTH 3.0V)
(Set current low limit 0A)
(Set current high limit 2A)
(Set NG Enable ON)
(Start OCP testing)
(Ask Testing? 1:Testing,0:Testing End)
(Ask PASS/FAIL?,0:PASS,1:FAIL)
(Ask OCP current value)
(Stop OCP testing)