BATT CF: setting the PF, the Left 5 digit monitor display the "BATT" ,the right
upper 5 digit monitor display the "PF", and right lower monitor display setting value.
The range is 0.01 ~1.00.
BATT TIME: setting the Battery test time, the Left 5 digit monitor display the
"BATT" ,the right upper 5 digit monitor display the "TIME", and right lower monitor
display setting value, the unit is "S". The range is 1S ~99999S.
BATT VTH: the Left 5 digit monitor display the "BATT" ,the right upper 5 digit
monitor display the "VTH", and right lower monitor display setting value, the unit is
"V". The range is 0.01V to the full scale of the Voltage specification.
• The TRANS parameters setting:
Pressing the Item key once will cause the button to illuminate. The message
"TRANS PRESS START" will be shown across the displays.
Each press of the setting button moves the menu on one step. The Left and right
LCDs show the currently selected test parameter as text. The value is adjusted by
The rotary knob and can be read from the Right display during Setting.
The setting sequence is shown below:
TRANS PRESS START
TRANS CC
TRANS CC : setting the Battery current point, the Left 5 digit monitor display the
"TRANS" ,the right upper 5 digit monitor display the "CC", and right lower monitor
display setting value, the unit is "A". The range is 0.001A to the full scale of the CC
mode specification.