Table 4-10 Register Of Prot Status - Prodigit 3270 Series Operation Manual

Ac/dc high power electronic load
Table of Contents

Advertisement

114 PRODIGIT
BIT ID
bit 0
bit 1
bit 2
bit 3
[STATe:] NGEABLE {ONOFF}
Syntax: [STATe:] NGEABLE {ONOFF} {;NL}
Purpose:To set the GO/NG check function enable or disable.
Description:To set the function of NG judgment opens when POWER ON. When setting
[STATe:] START
Syntax: [STATe:]START {;NL}
Purpose:Set for Load to implement the test.
Description:Set for Load to implement the test, and according to TEST CONFIG
[STATe:] STOP
Syntax: [STATe:] STOP {;NL}
Purpose:Set for Load to stop the test
Description:Set for Load to stop the test
[STATe:] TESTING?
Syntax: [STATe:]TESTING {?} {;NL}
Purpose:Check whether the current electronic load is in the test state.
Description:Check whether the current electronic load is in the test state, 1: testing 0:
Example: START
[STATe:] SYNCronize
Syntax: [STATe:]SYNCronize {SP}{ONOFF} {;NL}
Purpose:Load sync signal.
Description:Electronic load sync signal,1: SYNC ON 0: SYNC OFF。
4.7.4. SYSTEM Set and Read the Status of 3270 Series
[SYStem:] RECall{ SP }m{ ,n }
Syntax:[ SYStem:] RECall{ SP }m{;NL}
BIT VALUE
0 = Off, 1 = Triggered
0 = Off, 1 = Triggered
0 = Off, 1 = Triggered
0 = Off, 1 = Triggered

Table 4-10 register of PROT status

for POWER OFF, the function of NG judgment will not be implemented.
(TCONFIG), the Load will start to test the items and parameters which are
Required
Test end.
TESTING?
NG?
STOP
[STATe:]SYNCronize {?} {;NL}
REMARK
Over Power Protection (OPP)
Over Temperature Protection (OTP)
Over Voltage Protection (OVP)
Over Current Protection (OCP)

Advertisement