Table 12-1. Rear Panel Fan Assembly
Ref.
P Part
Qty
Desig.
Number
1
08 60-00016
2
1520-02 0
08 60-80006
4
5
08 60-20109
6
7
0515-0 72
8
08 60-6012
B1 Fan Assembly Removal
WARNING
These servicing instructions are for use by qualified personnel
only. To avoid electrical shock, do not perform any servicing
unless you are qualified to do so.
The opening of covers or removal of parts is likely to expose
dangerous voltages. Disconnect the instrument from all
voltage sources while it is being opened.
The detachable power cord is the instrument disconnecting
device. It disconnects the mains circuits from the mains
supply before other parts of the instrument. The front panel
switch is only a standby switch and is not a LINE switch.
1. Remove the two fan cage screws (7).
2. Remove the fan cage (6), lter (5), and lter retainer (4).
3. Remove the four fan screws (9).
4. Slide the fan out of the opening in the rear panel.
5. Disconnect the two-wire fan harness.
HP 8360
ervice
Description
2
FAN MOUNT
4
SHOCK MOUNT .27-EFF-HGT 2-LB-LOAD-CAP
1
FAN (B1)
1
FILTER RETAINER (Not available separately)
1
FOAM FILTER
1
FAN CAGE (Not available separately)
2
SCREW-MACHINE ASSY M X 0.5 8MM-LG
1
KIT - FAN FILTER ASSEMBLY (Includes fan cage, foam
lter, and lter retainer)
Change D 12-13