Calibration Constant Descriptions - Agilent Technologies 83620A Service Manual

Synthesized sweepers
Hide thumbs Also See for 83620A:
Table of Contents

Advertisement

Table 3-2. Calibration Constant Descriptions
Number
Description
105
SRD Bias A; Band 1
106
SRD Bias A; Band 2
107
SRD Bias A; Band
108
SRD Bias A; Band 4
109
SRD Bias A; Band 5
110
SRD Bias A; Band 6
111
SRD Bias A; Band 7
118
SRD Bias B; Band 1
119
SRD Bias B; Band 2
120
SRD Bias B; Band
121
SRD Bias B; Band 4
122
SRD Bias B; Band 5
12
SRD Bias B; Band 6
124
SRD Bias B; Band 7
1 1
Squegg Clamp 1A
1 2
Squegg Clamp 2A
1
Squegg Clamp A
1 4
Squegg Clamp 4A
1 5
Squegg Clamp 5A
1 6
Squegg Clamp 6A
1 7
Squegg Clamp 7A
144
YTM Delay Term A Hrm 1
145
YTM Delay Term A Hrm 2
146
YTM Delay Term A Hrm
147
YTM Delay Term A Hrm 4
148
YTM Delay Term A Hrm 5
149
YTM Delay Term A Hrm 6
150
YTM Delay Term A Hrm 7
3-36 Adjustments
Amplifier/Multiplier Adjustments
All manuals and user guides at all-guides.com
Default
Value
255
Maximize power over lower 0%
of band. Sweep speed is not a
1 0
factor. Adjust for 0.5 dB
1 0
below maximum power.
1 0
1 0
120
120
255
Maximize power over higher 0%
of band. Sweep speed is not a
150
factor. Adjust for 0.5 dB below
150
maximum power.
150
160
160
160
255
Default values are not altered.
Numbers should remain
255
unchanged (#1 1 is adjusted
255
in another procedure).
255
255
255
255
1000
Default values are not altered.
Numbers should remain
1000
unchanged. Maximize power over
1000
rst 20% of band. A ects
fast sweeps only. Single sweep
1000
mode is also critical.
1000
1000
1000
Adjustment
Description
HP 8360

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

83622a83623a83624a83630a83640a83650a

Table of Contents