Maximum Leveled Power - Agilent Technologies 83620A Service Manual

Synthesized sweepers
Hide thumbs Also See for 83620A:
Table of Contents

Advertisement

6. Maximum Leveled Power

his performance test is not valid unless the synthesizer meets
Note
both its power accuracy and power atness speci cations.
Perform those tests rst.
Description and Procedure
he unleveled status indicator is displayed when the instrument is unleveled
as the synthesizer sweeps over speci c frequency ranges in fast continuous
sweep, and fast and slow single sweep operation. Because of the synthesizer's
power accuracy and atness performance, a power meter is not required for
this measurement. he following procedure tests the most likely worst case
situations for maximum leveled power.
1. Set up and turn on the equipment shown in Figure 2-7. Preset the
synthesizer and let it warm up for at least one hour.
he 20 dB attenuator provides a good match on the RF
Note
output. If the synthesizer has a step attenuator, you can
simulate a good match by decoupling the step attenuator from
the ALC and setting the attenuator to 20 dB.
Figure 2-7. Maximum Leveled Power Test etup
Attenuator 20 dB
Attenuator 20 dB
2-34 Performance Tests
Maximum Leveled Power
All manuals and user guides at all-guides.com
Equipment
HP 8490D Option 020 (HP 83640A/50A)
HP 8493C Option 020 (HP 83620A/22A/23A/24A/30A)
HP 8360