All manuals and user guides at all-guides.com

Accuracy Versus Adjustment Interval

Figure 3-3 shows the required adjustment interval to maintain a given accuracy.
If you know the aging rate, you can determine a more precise adjustment
interval.
Figure 3-3. Accuracy Versus Adjustment Interval
1. Find the line on Figure 3-3 that corresponds to the 10 MHz timebase
oscillator aging rate. ( o determine the aging rate, see \Performance
ests".)
2. On the horizontal axis, nd the maximum allowable error (accuracy) that
you want.
3. Follow the maximum allowable error vertically until it intersects the known
aging rate.
HP 8360
Adjustments
3-9