Procedure - Agilent Technologies 83620A Service Manual

Synthesized sweepers
Hide thumbs Also See for 83620A:
Table of Contents

Advertisement

Procedure

1. Connect the equipment as shown in Figure 3-18 using an analog
oscilloscope with A versus B sweep capability. Do not connect the power
meter yet. (See Figure 3-19 for location detail of XA10J2 pin 55 on the
motherboard.) Power on all the instruments and let them warm up for at
least one hour.
2. Press
on the synthesizer and, for the HP 83640A/50A press:
PRESET
5
Tools Menu Disable Doubler
SERVICE
5
3. On the power meter:
Zero and calibrate the power meter/sensor.
Set the power meter to dBm mode.
Figure 3-18. Amplifier/Filter Adjustments etup
3-54 Adjustments
Amplifier/Filter Adjustments
All manuals and user guides at all-guides.com
(asterisk on)
HP 8360

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

83622a83623a83624a83630a83640a83650a

Table of Contents