Synthesizer Frequencies - Agilent Technologies 83620A Service Manual

Synthesized sweepers
Hide thumbs Also See for 83620A:
Table of Contents

Advertisement

All manuals and user guides at all-guides.com
Spectrum Analyzer
Function Generator
2. o achieve peak power on the synthesizer, turn on RF peaking. Press:
Tracking Menu Peak RF Always
USER CAL
5
3. On the synthesizer, press:
First synthesizer frequency in able 2-27
CW
5
Set the speci ed maximum leveled power
POWER LEVEL
5

Synthesizer Frequencies

1
Disregard if beyond your synthesizer's capability.
4. On the function generator, set:
Function:
ine wave
Frequency: 1 MHz
Amplitude: 480.8 mV p-p (169.96 mV rms)
DC Offset: 0 V
5. On the spectrum analyzer, set:
Reference Level:
Attenuator:
Center Frequency:
Frequency pan:
HP 8360
Equipment
Table 2-27. ynthesizer Frequencies
External FM
(GHz)
Error (%)
1.0
1
5.0
10.0
16.0
22.0
1
0 dBm
Auto
ame frequency as in step 3
50 kHz
HP 8566B
HP 3325A
(asterisk on)
Internal FM
Error (%)
Performance Tests
FM Accuracy
2-115

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

83622a83623a83624a83630a83640a83650a

Table of Contents