21. FM Input Impedance
Description and Procedure
A soldering iron is required for this procedure.
his procedure sets the FM input impedance to either 50
or 600
. he
FM input impedance on A11 must be set to 600
for proper operation of the
internal modulation generator (Option 002).
Instruments without Option 002 should perform step A, \A11 Jumper".
Instruments with Option 002 should set the A11 jumper to the 600
position
using step A, \A11 Jumper" and should perform step B, \A8 Jumper".
HP 8360
Adjustments
3-117
FM Input Impedance