Setting A Password; Bypassing The Password - Agilent Technologies 83620A Service Manual

Synthesized sweepers
Hide thumbs Also See for 83620A:
Table of Contents

Advertisement

All manuals and user guides at all-guides.com

Setting a Password

If a password is already set on the synthesizer and you wish to change it, rst
follow the \Disabling a Password" procedure. his eliminates the current
password. hen continue with this procedure.
If no password is set on the synthesizer (there is no asterisk on the
softkey), and you wish to set a password, perform the
Disable Adjust
following:
1. On the synthesizer, set:
Disable Adjust
SERVICE
5
he following message is displayed:
2. Enter the desired password using the numeric keypad. Up to 14 numeric
digits are allowed. erminate the entry by pressing:
3. A warning message is displayed informing you that the adjustment menus
will no longer be accessible. o verify that you do wish to set the password,
select
Yes, Confirm
he asterisk on the
Disable Adjust
password is set.

Bypassing the Password

If you require access to the adjustment menu for calibration purposes, a
password is set, and you do not know the password, perform the following:
1. urn o line power to the synthesizer.
2. Open A15S1 switch 5 (set the switch to the
3. urn on line power to the synthesizer and press
he asterisk on the
the adjustment menu.
When access to the adjustment menu is no longer desired, enable the
previously set password. Perform the following:
4. urn o line power to the synthesizer.
HP 8360
.
SET P SSWORD: 0
.
softkey turns on, indicating the
softkey turns o enabling access to
Disable Adjust
.
ENTER
5
1
position).
"
"
.
SERVICE
5
Calibration Constants 4-5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

83622a83623a83624a83630a83640a83650a

Table of Contents