Post{Repair; Replacement Of An Assembly - Agilent Technologies 83620A Service Manual

Synthesized sweepers
Hide thumbs Also See for 83620A:
Table of Contents

Advertisement

All manuals and user guides at all-guides.com

Post{Repair

Adjustments and Performance Tests Required After Repair or
Assembly
A1 Front Panel Keyboard
A2 Source Module Interface
A Front Panel Processor
1
A4 Fractional{N
A5 YO Phase Detector
A6 Sampler
1
The adjustments listed are required only after
adjusted at the factory and is not instrument-dependent.
HP 8360
Table 8-1.

Replacement of an Assembly

Adjustment
None
None
None
Fractional{N VCO
Fractional{N Reference
and API Spurs
YO Loop Gain
Sampler Assembly
of the assembly. The
repair
8
Performance Test
Full Self{Test
Front Panel Self{Test
Full Self{Test
Full Self{Test
Front Panel Self{Test
Full Self{Test
Frequency Switching Time
Spurious Signals
(Non-Harmonics)
Full Self{Test
Frequency Switching Time
Spurious Signals
(Non-Harmonics)
Single Sideband Phase Noise
Full Self{Test
Frequency Switching Time
Spurious Signals
(Non-Harmonics)
Single Sideband Phase Noise
assembly is
repla ement
Post{Repair 8-1