Entering a Password

An asterisk on the
Disable Adjust
a password is set. In order to access the adjustment menu you must enter that
password:
1. On the synthesizer, set:
Adjust Menu
SERVICE
5
he following message is displayed:
2. Enter the correct password using the numeric keypad and press
he asterisk on the
Disable Adjust
3. Select
Adjust Menu
Entering the password allows permanent access to the adjustment menu, even
if line power is cycled. o disable access again, set another password.

Disabling a Password

If you wish to eliminate a password (either the factory-set password or one you
set yourself), perform the following:
1. On the synthesizer, set:
Disable Adjust
SERVICE
5
he following message is displayed:
2. Enter the current password using the numeric keypad and press
he asterisk on the
Disable Adjust
Disabling the password allows permanent access to the adjustment menu, even
if line power is cycled.
4-4 Calibration Constants
All manuals and user guides at all-guides.com
softkey in the service menu indicates that
.
ENTER P SSWORD: 0
softkey turns o .
again to access the adjustment menu.
.
ENTER P SSWORD: 0
softkey turns o .
.