2. Leaving the exible cable connected to A6J1 (40 MHz reference), place the
A6 assembly on the analog extender board.
3. urn the synthesizer on and press
hour.
4. Select the A6 adjustment menu. Press:
Adjust Menu AssyAdj Menu A6 Adj Menu
SERVICE
5
200 MHz Loop VCO Tune Check
5. In the A6 adjustment menu, select the A6 VCO tune check
6. he following message should be displayed:
Ctrim and C60 are OK.
If this message is not displayed, refer to troubleshooting.