Change the value of the calibration constant using the right/left arrow keys,
the rotary knob, or the numeric keypad. erminate numeric keypad entries
by pressing
.
ENTER
5
4. Exit the calibration menu using the
Saving Working Data Calibration Constants
In some adjustment procedures you will change working data calibration
constants. he following procedure stores the calibration constants as
protected data (in EEPROM).
1. On the synthesizer, set:
Adjust Menu Calib Menu Cal Util Menu
SERVICE
5
2. Select
. A warning is displayed informing you that changing the
ave Cal
calibration constants may drastically aect instrument performance. o
verify that you do wish to save the new calibration constants, select
he working data calibration constants are now stored as protected data
(in EEPROM).
3. Exit the calibration utility menu using the
Loading Protected Data Calibration Constants
If your working data calibration constants have been altered or deleted, the
following procedure loads the protected calibration constants from EEPROM
into working data memory.
1. On the synthesizer, set:
.
Cal Util Menu
2. Select
.
Recall Cal
HP 8360
key or by pressing
PRIOR
5
PRIOR
Adjust Menu Calib Menu
SERVICE
5
.
PRESET
5
.
yes
key or pressing
PRESET
5
Calibration Constants 4-7
.
.
5