13. Set the function generator to
on the DVM.
14. On the synthesizer, press:
AM On/Off 10dB/V
or, for synthesizers with Option 002, instead press:
(asterisk on)
AM On/Off Ext
(asterisk on)
AM Type 10dB/V
15. Adjust A10R35, EXP AM CAL, for a
Figure 3-29 for the location of A10R35.
Figure 3-29. A10R35 and A10R39 Locations
Linear AM Accuracy
16. On the synthesizer, press:
AM On/Off 10dB/V
or, for synthesizers with Option 002, instead press:
(asterisk o)
AM On/Off Ext
17. On the synthesizer, press:
POWER LEVEL
5
dBm
5
5
3-104 Adjustments
AM Accuracy
All manuals and user guides at all-guides.com
0
1.0 V DC oset. Verify the correct voltage
(asterisk on)
0
10 dB power meter reading. See
(asterisk o)
5
HP 8360