All manuals and user guides at all-guides.com
Millimeter Band Power Flatness
Disregar if beyon your synthesizer's capability.
9. Load and select the calibration factors for the correct power sensor for this
frequency band into the power meter. Refer to the power meter manual for
instructions.
Zero and calibrate the power meter then connect the power sensor to the
synthesizer RF output.
10. On the synthesizer, press:
Cal Freq Range mm Band
11. Repeat steps 3 through 5.
Related Performance Tests
Power Flatness
In Case of Difficulty
1. he HP 437B Power Meter must be used. Correct calibration factors must
be loaded and selected.
2. Make sure that the only HP-IB connection is between the synthesizer and
the power meter. No controller is allowed on the bus.
3. Refer to the
Troubleshooting
HP 8360
(asterisk on)
manual.
Adjustments
3-99
Power Flatness