A warning is displayed informing you that changing the calibration
constants may drastically a ect instrument performance. o verify that you
do wish to recall the protected data calibration constants, select
he protected data calibration constants are now loaded into working data.
3. Exit the calibration utility menu using the

Loading Default Data Calibration Constants

If you want to use the generic synthesizer calibration constants as working
data, the following procedure loads the default calibration constants from
UVEPROM into working data memory.
1. On the synthesizer, set:
Adjust Menu Calib Menu Cal Util Menu
SERVICE
5
2. Select
.
Default Cal
A warning is displayed informing you that changing the calibration
constants may drastically a ect instrument performance. o verify that you
do wish to recall the default calibration constants, select
he default data calibration constants are now loaded into working data.
3. Exit the calibration utility menu using the

Calibration Constant Descriptions

able 4-9 provides the following information for each calibration constant:
he calibration constant number in ascending order.
he calibration constant name. An asterisk indicates that this calibration
constant is adjusted by an automated adjustment.
he range of realistic adjustment (adjustment outside of the range may be
possible but will probably cause the synthesizer to be severely misadjusted)
4-8 Calibration Constants