A10 Jumper - Agilent Technologies 83620A Service Manual

Synthesized sweepers
Hide thumbs Also See for 83620A:
Table of Contents

Advertisement

All manuals and user guides at all-guides.com
A. A10 Jumper
1. On the synthesizer, set the line power switch to standby.
2. Lift up the A10 ALC assembly far enough to access the jumper shown in
Figure 3-35.
3. he AM input impedance is factory-set to the 50 ohm position. o change
it to 2 k , move the jumper to the position marked \2K". he jumper can
be returned to the \50 OHM" position to change the AM input impedance
to 50 ohms.
4. Reinstall the A10 assembly.
Figure 3-35. A10 Jumper Locations
HP 8360
Adjustments
3-115
AM Input Impedance

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

83622a83623a83624a83630a83640a83650a

Table of Contents