4. J1 RF Output Connector Removal (Option 004)

Refer to Figure 7-20.
o avoid damaging the connector, do not disassemble the RF
CAUTION
output connector assembly. Only unscrew the connector on the
end of the cable.
1. Complete \1. Preliminary Steps" before continuing.
2. Disconnect cable W43 at the directional coupler A30 (or at the step
attenuator A31 for instruments with Option 001).
3. Remove the two screws from the RF output connector plate on the rear
panel.
4. Remove the RF connector with cable W43 attached.
Figure 7-20. Removing RF Output Connector