1. Preliminary Steps
You can easily blow the fuse for the 13.5V supply on the
CAUTION
preregulator assembly if you do not disconnect the AC power
cord. Also, the danger of shock is increased if the power cord is
connected.
1. Disconnect the AC power line cord.
2. Remove the screw from each of the top two feet on the rear panel.
3. Remove the top two feet.
4. Remove the instrument top cover.
An RF braid is in each of the side channels on the top of the
Note
instrument. When you reassemble the instrument, be sure this
braid remains in the channel.
5. Remove the vinyl trim strip from the top front edge of the instrument.
(Place a
at screwdriver in either slot in the trim strip and lift to remove,
see Figure 7-3.)
7-6 Disassembly and Replacement Procedures