13. Pulse Performance

Description and Procedure
In this procedure, the synthesizer's RF output is displayed directly on a
high-bandwidth digitizing oscilloscope where the rise and fall times are
measured and a comparison of pulsed and CW power level accuracy is made.
Power level accuracy is checked at a 10 Hz pulse repetition rate to verify its
operation at the slowest speci ed rate.
Rise and Fall Times
1. Connect the equipment as shown in Figure 2-15. Preset the instruments
and let them warm up for at least one hour.
Figure 2-15. Pulse Performance Test etup
2-68 Performance Tests
HP 8360
Pulse Performance