2. YO Delay Adjustment
Description
his adjustment uses the internal counter to measure swept frequency
accuracy. YO delay aects the swept frequency accuracy at fast sweep times
(less than 100 ms). his procedure adjusts the calibration constants for the
YIG oscillator to achieve consistent swept frequency accuracy over the entire
operating range of the instrument.
If your instrument has Option 006, be certain that you
CAUTION
identied your option conguration when the software was
initially loaded or else this procedure will not run properly.
Select the YO delay adjustment and follow the prompts on the display.
No test equipment is required.