Procedure
In this procedure, some calibration constants and
Note
potentiometers may be set to default conditions and left. his
is normal. hese boards contain adjustments for dierent
vintages of the HP 8360 family. Not all adjustments are used
each time.
1. Connect the equipment as shown in Figure 3-14 using an analog
oscilloscope with A versus B sweep capability. Power on all the instruments
and let them warm up for at least one hour.
2. On the synthesizer, press
Figure 3-14. Amplifier/Multiplier Adjustments etup
3-34 Adjustments
Amplifier/Multiplier Adjustments
All manuals and user guides at all-guides.com
.
PRESET
5
HP 8360