High Band Power Flatness; Millimeter Band Power Flatness 20 Ghz And 40 Ghz; Millimeter Band Power Flatness 40 Ghz - Agilent Technologies 83620A Service Manual

Synthesized sweepers
Hide thumbs Also See for 83620A:
Table of Contents

Advertisement

All manuals and user guides at all-guides.com

High Band Power Flatness

7. With the appropriate power sensor for this frequency range, repeat steps 2
through 6 for:
A 20 GHz stop frequency.
A 2.05 GHz start frequency.
A 100 MHz increment.
Millimeter Band Power Flatness
Disregar if beyon your synthesizer's capability.
8. With the appropriate power sensor for this frequency range, repeat steps 2
through 6 for:
A 40 GHz stop frequency (for HP 83630A only, set to 26.5 GHz).
A 20.05 GHz start frequency.
A 100 MHz increment.
Millimeter Band Power Flatness
Disregar if beyon your synthesizer's capability.
9. With the appropriate power sensor for this frequency range, repeat steps 2
through 6 for:
A 50 GHz stop frequency.
A 40.05 GHz start frequency.
A 100 MHz increment.
Related Adjustments
Power Flatness
In Case of Difficulty
1. he power meter must be an HP 437B.
2. he correct calibration factors for the power sensor must be loaded.
3. Refer to the
Troubleshooting
HP 8360

20 GHz and
40 GHz
40 GHz
manual.