All manuals and user guides at all-guides.com
High Band Power Flatness
7. With the appropriate power sensor for this frequency range, repeat steps 2
through 6 for:
A 20 GHz stop frequency.
A 2.05 GHz start frequency.
A 100 MHz increment.
Millimeter Band Power Flatness
Disregar if beyon your synthesizer's capability.
8. With the appropriate power sensor for this frequency range, repeat steps 2
through 6 for:
A 40 GHz stop frequency (for HP 83630A only, set to 26.5 GHz).
A 20.05 GHz start frequency.
A 100 MHz increment.
Millimeter Band Power Flatness
Disregar if beyon your synthesizer's capability.
9. With the appropriate power sensor for this frequency range, repeat steps 2
through 6 for:
A 50 GHz stop frequency.
A 40.05 GHz start frequency.
A 100 MHz increment.
Related Adjustments
Power Flatness
In Case of Difficulty
1. he power meter must be an HP 437B.
2. he correct calibration factors for the power sensor must be loaded.
3. Refer to the
Troubleshooting
HP 8360
20 GHz and
40 GHz
40 GHz
manual.
Performance Tests
2-33
Power Flatness