10. Select
.
Done

200 MHz Loop Gain Adjustment

11. In the A6 adjustment menu, select the A6 loop gain adjustment,
. he synthesizer automatically performs the adjustment.
A6 Loop Gain
When the message
Calibration Constants should be S VED
displayed, the adjustment is complete. Continue with the next step.
12. Set the synthesizer to standby. Reinstall the A6 assembly in the
instrument, and connect all cables.

IF Gain Adjustment

13. urn the synthesizer on. Select
14. Adjust A6R73 (see Figure 3-9) to center the \needle" on the display.
Figure 3-9. IF Gain Adjustment Location
15. Select
.
Done
3-22 Adjustments
ampler Assembly
All manuals and user guides at all-guides.com
A6 IF Gain