Yo Gain And Linearity Adjustment Locations - Agilent Technologies 83620A Service Manual

Synthesized sweepers
Hide thumbs Also See for 83620A:
Table of Contents

Advertisement

All manuals and user guides at all-guides.com
If the following message is displayed:
Note
until "locked".
indication is displayed. Continue with step 3.
3. Select the gain potentiometer adjustment,
4. Adjust A13R54 (see Figure 3-12) to center the \needle" on the display.
Figure 3-12. YO Gain and Linearity Adjustment Locations
5. Select the low break point potentiometer adjustment,
6. Adjust A13R90 (see Figure 3-12) to center the \needle" on the display.
7. Select the high break point potentiometer adjustment,
8. Adjust A13R92 (see Figure 3-12) to center the \needle" on the display.
9. Select
.
Done
10. If this is the last calibration constant you will be adjusting, see
\Calibration Constants" to store the calibration constants as protected
data (in EEPROM).
HP 8360
djust 13 pot
, adjust the potentiometer until the locked
.
Gain Pot
YO Driver Gain and Linearity
.
Lo Bk Pot
.
Hi Bk Pot
Adjustments
3-27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

83622a83623a83624a83630a83640a83650a