Fractional-N Vco; A4L1 Location - Agilent Technologies 83620A Service Manual

Synthesized sweepers
Hide thumbs Also See for 83620A:
Table of Contents

Advertisement

All manuals and user guides at all-guides.com
2. Fractional- VCO
Description and Procedure
No test equipment is required for this procedure.
his adjustment sets the VCO tuning voltage to
60 MHz.
1. Leaving all cables connected, place the A4 fractional-N assembly on an
analog extender board.
2. urn the synthesizer on and press
at least one hour.
3. Select the A4 VCO tune adjustment. On the synthesizer, press:
Adjust Menu AssyAdj Menu A4 VCO Tune
SERVICE
5
4. Using a non-metallic tool, adjust A4L1 (see Figure 3-4) to center the needle
on the display.
5. urn the synthesizer to standby, and reinstall the A4 assembly.
HP 8360
0
. Let the instrument warm up for
PRESET
5
Figure 3-4. A4L1 Location
6.0 V for a VCO output of
Adjustments

Fractional-N VCO

3-11

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

83622a83623a83624a83630a83640a83650a

Table of Contents