12. Modulator Offset and Gain
Description and Procedure
In this procedure, the ALC modulation oset and gain calibration constants
are adjusted to linearize the ALC modulator response to the ALC power
level reference voltage. Default values are entered for the modulator oset
calibration constants and internal rmware is activated to set the modulator
gain calibration constants. he synthesizer is then set for a power sweep across
the entire leveled ALC range (
integrator level signal on the ALC board is monitored to verify linearity.
If necessary, the modulator oset values are modied.
1. Connect the equipment as shown in Figure 3-23. Preset the instruments
and let them warm up for at least one hour.
Figure 3-23. Modulator Offset and Gain Adjustment etup
Digitizing Oscilloscope
Oscilloscope Probes
3-86 Adjustments
Modulator Offset and Gain
All manuals and user guides at all-guides.com
0
20 to the maximum leveled power). he
Equipment
HP 54111D
HP 10431A
HP 8360