Figure 3-5. Fractional-N Reference and API pur Adjustment etup
Spectrum Analyzer
2. On the spectrum analyzer, set:
Frequency pan:
Resolution Bandwidth: 10 Hz
weep Time:
Reference Level:
cale Log:
Video Averaging:
Clear Write A:
125 kHz Reference pur Adjustment
3. On the synthesizer, press:
CW
5
Tools Menu CntlPLL Menu Frac N Menu
SERVICE
5
CW
44.125
MHz
5
5
5
3-14 Adjustments
Fractional-N Reference and API purs
All manuals and user guides at all-guides.com
Equipment
0 Hz
10 s
0
10 dBm
10 dB/Division
Off
elected
HP 8566B
HP 8360