Turn off power before replacing the fuse.
Fuse specifications
AC power source (100 or 200 V): 5.0 mm (diameter)  20 mm (long), 250 V, 0.5 A.
(1)
Clearing characteristics (T), breaking capacity (L)
As represented by Littelfuse 0218.500MXP 250 V, 0.5 A.
DC power source: 5.2 mm (diameter)  20 mm (long), 250 V, 1 A.
(2)
Clearing characteristics (T), breaking capacity (L)
Example:0218001.MXP 250V,1A manufactured by littelfuse.
(1) Opening the cover after turning off power
Loosen 4 screws from the flow transmitter front, and open the cover.
(2) Replacing the fuse
Detach the fuse holder from the power supply board, and replace the fuse. Then, return the fuse holder in place.
(3) Closing the cover
Close the cover, and tighten 4 screws.
Turn on power only after closing the cover.