10
Hardware Information

Setting the 8-bit Configuration Switch

Setting the 8-bit Configuration Switch

The 8- bit configuration switch is located at the rear of the module. Switch
settings provide configuration parameters for LAN, serial communication
protocol and instrument specific initialization procedures.
All modules with on- board LAN:
• Default is ALL switches DOWN (best settings).
• For specific LAN modes switches 3- 8 must be set as required.
• For boot/test modes switches 1+2 must be UP plus required mode.
For normal operation use the default (best) settings.
N O T E
Figure 22
184
Bootp mode for LAN and
19200 baud, 8 data bit / 1 stop bit with no parity for RS- 232
Location of Configuration Switch (example shows a G4212A DAD)
Agilent 1290 Infinity II Flexible Pump User Manual