Setting the 8-bit Configuration Switch

The 8-bit configuration switch is located next to the GPIB connector. Switch
settings provide configuration parameters for GPIB address, serial
communication protocol and instrument specific initialization procedures.
Figure 48
Table 54
Mode Select
GPIB
RS-232C
Reserved
TEST/BOOT
Switches 1 and 2 define which set of parameters (for example, for GPIB,
RS-232C, and so on) will be changed. Once the change has been completed, the
instrument must be powered up again in order to store the values in the
non-volatile memory.
1100 Series Nano Pump Service Manual
8-bit Configuration Switch
8-bit Configuration Switch
1
2
3
0
0
0
1
Baudrate
1
0
Reserved
1
1
RSVD
Introduction to the Nano Pump
4
5
6
GPIB Address
Data
Bits
SYS
RSVD
5
7
8
Parity
RSVD
FC
223