7
Maintenance
Install the Jet Weaver
3
Insert the Jet Weaver to the opening in the front panel (1)
and push it down (2).
5
Mount both capillary connections to the Jet Weaver
observing the correct orientation.
7
Configure the Jet Weaver as mixer in the user interface.
98
4
Fix the Jet Weaver with a screw.
6
Connect the inlet capillary of the Jet Weaver to port 2 of
the Multi Purpose Valve. Connect the outlet capillary to
port 1.