Drive current too low

Error ID: 29205
The current consumption of the pump drive is too low.
• Parameter: 1 – 2 referring to pump drive
Probable cause
1
2

Drive current too high

Error ID: 29236
The current consumption of the pump drive is too high.
• Parameter: 1 – 2 referring to pump drive
Probable cause
1
2
Agilent 1290 Infinity II Flexible Pump User Manual
Drive motor defect.
Wrong/missing connection of pump drive to
main board.
Blockage of system before pressure sensor.
Drive motor defect.
Error Information
Pump Error Messages
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Check for blockage of e.g. outlet valve filter frit,
Multi Purpose Valve, heat exchanger.
Please contact your Agilent service
representative.