6
Error Information
Pump Error Messages

Valve hardware overcurrent (MCGV)

Error ID: 29227
Power consumption too high for one of the MCGV valves.
Probable cause
1
2
3
78
Cable defect.
Valve defect
Defective main board.
Suggested actions
Replace MCGV.
Replace MCGV.
Please contact your Agilent service
representative.